All Categories

Welcome To


                          ORNET

HR2000+CG SPECTROMETER

Description

The HR2000+CG spectrometer is perfect for applications where fast measurements and high resolution are needed.

 

Specification Of HR2000+CG Spectrometer

Engineering Specifications

HR2000+CG

PHYSICAL

 

Dimensions: (L x W x H) mm and inches

148.6 x 104.8  x 45.1 mm (5.9 x 4.1 x 1.8 in.)

Weight: kg and lb

0.57 kg (1.26 lbs.)

DETECTOR

 

Type:

Sony ILX511B

Range:

190 – 1100 nm

Quantum efficiency:

no

SPECTROSCOPIC

 

Wavelength range:

200-1050 nm

Integration time:

1 millisecond – 65 sec

Dynamic range:

8.5 x 107 (system); 1300:1 for a single acquisition

Signal-to-noise ratio:

250:1 (at full signal)

Grating:

HC-1

Slit:

5 µm wide

Optical resolution:

1.0 nm FWHM

Stray light:

<0.05% at 600 nm; <0.10% at 435 nm

Buffering:

no

Fiber optic connector:

SMA 905 to single-strand optical fiber (0.22 NA)

ELECTRONICS

 

Power consumption:

220 mA at +5 VDC

Strobe functions:

2 programmable strobe signals (single/continuous)

Interfaces:

USB 2.0, 480 Mbps; 2-wire RS-232; I2C 2-wire serial bus

Temperature:

-30° to +70° C Storage & -10° to +50° C Operation

Humidity:

0% – 90% non-condensing

 

Applications in which this product is used:

Protein & Nucleic Acid Analysis
Food & Beverage Quality Control
Remote Sensing