ORNET

Single Flame Atomic Absorption Spectrophotometer

FAAS Spectrophotometer

The Single Flame Atomic Absorption Spectrophotometer (FAAS) AS2000 is a brand new product carefully crafted by Optosky. The original optical path design greatly improves the sensitivity in the ultraviolet region. The outstanding mechanical and optical system, coupled with excellent software, ensure the accuracy of measurement.

The economical single flame atomic absorption spectrophotometer AS2000 can install 4 hollow cathode lamps at the same time, providing 6 slits (automatic switching); the instrument has good stability, wide measurement range, low detection limit, high resolution and high sensitivity; The software adds a method storage function. Users can store the standard curve and call it up at any time, which can save analysis time and adapt to your pursuit of automated and accurate measurement results.

 

Portable Hyperspectral Imgaer Get Best Price

SPECIFICATION

Lo

Model

AS2000

Wavelength Range

190~900nm

Grating Line Density

1800 lines /mm

Absorbance Range

0–2.5ABS

Wavelength Repeatability

±0.1nm

Wavelength Accuracy

Full spectrum ±0.2 nm

Optical System

Integrated optical platform, fully enclosed optical system (single beam)

Detector

Photomultiplier tube

Resolution

Able to separate manganese doublet lines (279.5 and 279.8 nm) with a spectral bandwidth of 0.2 nm and a valley-to-peak energy ratio of <30%

Spectral Slits

0.1nm,0.2nm,0.4nm,1nm,1.2nm.2.0nm(automatically switchable)

Background Correction Technology

D2 background subtraction (background signal 1 ABS, background reduction capability ≥50 times)

Static Baseline Drift

≤0.004ABS/30min( Cu)

Lamp Holder

4-lamp turret

Characteristic Concentration (Copper)

≤0.025ug/ml

Measurement Repeatability(Copper)

≤0.5%(Cu,Absorbance>0.8ABS)

Detection Limit

Cd ≤0.3pg

Gas Flow Control

Needle valve control

Safety Protection

Automatic gas cut-off for low pressure, power interruption, abnormal flameout, and burner mismatch;

Background Subtraction Method

Self-absorption and deuterium lamp background subtraction methods,can correct 1A background. When the background signal is 1A, the background deduction ability is ≥50 times

Burner

Air-acetylene flame burner 100mm

Ignition Method

Automatic ignition

Ignition Dynamic Baseline Drift

≤0.004ABS/30min(Cu)

Dimensions

700*500*530(mm)

Weight

80kg

Feature

  • HCL:4 hollow cathode lamps (can be installed)
  • Safety control for abnormal situations, such as insufficient pressure, air leakage, flameout.
  • Provides 6 levels of slit (automatic switching)
  • The software adds a method storage function. Users can store the standard curve and call it up at any time, which can save analysis time.
  • Good stability, wide measurement range, low detection limit

Application

  • Minerals:rare earth analysis and precious metals analysis
  • Environmental:Environmental Protection &Water quality testing
  • Material science:alloy materials & building materials
  • Scientific research: Elemental quantitative analysis in scientific research such as medical and health, and higher education institutions