GL OPTI SPHERE 2000
Large Integrating Sphere
LM79 compliant luminous flux and radiant power measurement of large luminaires.
GL OPTI SPHERE 2000 Usage
Measure. Launch. Grow.
With more and more companies manufacturing commercial lighting fixtures, there is increasing demand for a reasonably priced, accurate and easy to use solution to test & measure large luminaires during development and as a QC measure in production. GL OPTI SPHERE 1500, 2000 and 3000 integrating spheres were developed to meet this need and offer a completely turnkey solution for all your light measurement needs, big or small.
GL large integrating spheres deliver exceptional luminous flux and color accuracy without the operational complexity associated with other photometric test equipment. A dedicated metrologist or large dark lab are not required. Set up the system and start evaluating your lamps in no time. Our customers appreciate that you can just open the sphere, install the lamp inside and measure. This sphere is equipped with all accessories and components you need and our GL SPECTROSOFT software interface will help you to do it right.
A laboratory of your own
This instrument is equally suited to accredited laboratory environments, production labs or in R&D for quick quality control. When using external labs becomes too expensive or time-consuming, the GL LARGE OPTI SPHERE is an excellent choice for LM79 light measurement applications.
Instant customized reports
Install, measure and prepare customized report directly from our GL SPECTROSOFT suite. This analytical software will control the measurement instrument, auxiliary light source and optionally manage peripheral devices like programmable and stabilized power supplies or TEC controllers. All these can be included in one complete measurement report.
APPLICATION
- Luminous flux and radiant power measurement of large LED modules and large luminaires
GL OPTI SPHERE 2000 Specifications
| TECHNICAL DATA | |
| Spectral range* | 340 – 1700 nm |
| Sphere inner diameter | 2000 mm |
| Entrance aperture diameter | 660 mm |
| Sphere material | Carbon steel |
| Inner coating | Barium Sulfate (BaSO4) high-reflectance material (R98) |
| Outer coating | Black textured finish |
| Reflectance properties | 97% |
| Auxiliary light source | White LED or halogen |
| Spectrometer port | Direct connection or SMA fiber-optic |
| Standards compliance | CE, LM 79, CIE 127:2007, CIE S 025/E:2015 |
| Maximum DUT dimensions in accordance with CIE S 025/E:2015 | 200 mm (diameter or diagonal) |
| Maximum dimension for optimal measurement (1/3 x sphere diameter) | 665 mm (diameter or diagonal) |
| Maximum DUT weight | 25 kg |
| Sphere frame | Hinged |
| Sphere center positioning | Cross laser mechanism |
| Mechanical breadboard with post | For 4π measurement |
| USB source controller for auxiliary light source | With current source and relay switch for external power supply |
| Universal post with standard lamp sockets | Universal DUT fixing table (breadboard) for measurement in 4π geometry |
| External dimensions [W x H x D] | 2200 mm x 2200 mm x 2300 mm |
| Weight | 420 kg |
* Spectral range of the sensor. Actual spectral range of system may be reduced due to limitations of used optical accessory.