Infrared Multi-Gas Analyzer GF330

Gas Analyzer

Advanced multi-gas sensing solutions that can sense a variety of gases at a time. Explosion-proof products adapt to complex working conditions.

GF330 Infrared Multi-Gas Analyzer

Description

The infrared gas analyzer adopts non-dispersive infrared technology (NDIR, Non-Dispersive InfraRed). NDIR is a method based on the Lambert-Beer law gas absorption theory. After the infrared radiation emitted by the infrared light source is absorbed by a certain concentration of gas, the spectral intensity proportional to the gas concentration will change, so the concentration of the gas to be measured can be reversed by calculating the change in spectral light intensity.

Another core technology of infrared gas analyzer is based on infrared correlation filtering technology and self-designed long optical path gas absorption cell to realize quantitative analysis of gas in infrared band, which has the characteristics of high precision, good stability and fast response time.

     FEATURES

  • Continuous measurement of SO2, NO, CO, CO2, CH4, N2O, O2, SF6 and other gas concentrations
  • Support 4-20mA analog output, RS232, RS485, network port and other data output methods
  • Easy to operate, quick to set up
  • Color large-screen LCD operation interface, supports switching between Chinese and English
  • Directly complete parameter setting, measurement and calibration through the menu
  • Measuring range can be customized

     APPLICATION

  • Scientific research fields
  • Agricultural field
  • Industrial field
  • Environmental protection field
  • Security field


Parameter

Measuring principleInfrared absorption (NDIR)
Measuring gasCO, CO2, O2, etc. (expandable: SO2, NO, N2O, CH4, SF6; replace the corresponding sensor module according to customer needs)
Measuring rangeSO2: 0-100ppm to 0-2000ppm
NO: 0-1000ppm to 0-30000ppm
CO: 0-100ppm to 0-100% Vol
CO2: 0-100ppm to 0-100% Vol
CH4: 0-50ppm to 0-5% Vol
SF6: 0-50ppm to 0-1000ppm
N2O: 0-50ppm to 0-500ppm
O2: 0-5% Vol to 0-25% Vol
Preheat time5 minutes (depending on the sensor)
Response time≤5s
StabilityZero drift: ≤1% FS/week
Range drift: ≤1% FS/week
Output interfaceNumerical output: RS232/RS485
Switch output: IN/OUT (supports 8 channels)
Analog output: 4-20mA (supports 4 channels)
Gas inlet flow0.5L-1L/min
Linearity error≤1% FS
Repeatability≤1%
Ambient temperature effectWhen the ambient temperature changes by 10 °C, zero point and span change ≤2%
Atmospheric pressure effectPer change ≤1%, (allowable atmospheric pressure range) ≤2%
Use environmentTemperature: 0-45 °C
Humidity: 0-90%RH
Atmospheric pressure: 700-1200 hPa
Power supplyAC220V
Dimensions19 inch 4U sheet metal chassis
Weight12 kg